At a glance
| Term on a data sheet | What it identifies |
|---|---|
| Grade | Specified material composition or grade designation |
| Cold working | Processing that can change mechanical properties |
| Tensile strength | Stress measured under a defined pulling test |
| Finished-device performance | Depends on geometry and loading as well as material |
Reading a materials table
A study of commercially pure titanium and titanium alloy tested bars, discs and finished implants made from several grades. It included a modified, hardened grade 4 alongside conventional grade 4. The investigators performed different tests for tension, compression, hardness and torque. Tensile strength describes a specimen pulled in a particular way. Hardness measures a different response at its surface.

One modified grade 4 in one study
The hardened grade 4 showed higher measured mechanical strength than the conventional grade 4 and the grade 5 alloy in that experiment. That comparison applies to the materials and manufacturing conditions tested. It does not mean any grade 4 product has greater strength than every grade 5 product. The clinical performance comparison reported by the authors was similar for modified and conventional grade 4.
From a material bar to the planned implant
A bar tested in tension has a different shape and loading pattern from a threaded fixture attached to an abutment. For a narrow implant, the exact dimensions and how the internal connection occupies that space also matter to the assembly tested. Ask for the alloy or grade, the named product and the kind of test behind a strength claim. That makes a number in megapascals traceable to the material and specimen that produced it.