At a glance
| Record | Meaning in the clinical comparison |
|---|---|
| Sample | 13 arches in 11 people |
| Supports | Four implants per arch |
| Reference | Conventional splinted impression |
| Outcomes | Linear position in micrometers and orientation in degrees |
Position and angle answer different questions
A linear deviation describes how far one recorded position lies from the position in a reference scan. An angular deviation describes a difference in orientation. A system can reproduce a point closely and still record the direction imperfectly. Comparisons should keep micrometers and degrees in their original units, and name the reference method used to calculate both.


Two ways to record four implants
A prospective clinical study compared a conventional intraoral scanner and an intraoral photogrammetry scanner against a conventional splinted open-tray impression. Eleven people contributed 13 arches, each with four implants. Mean positional deviations were 59.8 micrometers for the conventional scanner and 30.7 micrometers for the photogrammetry system. The corresponding mean angular deviations were 1.4 degrees and 0.78 degrees.
What still needs to happen after a scan
These results describe agreement with the reference in the equipment and arches studied. They do not establish that every scan of a particular arch is within that distance or predict the lifetime of the final restoration. Ask the provider how they verify the actual framework or bridge when it is fitted, and which implant-level or abutment-level components were recorded. A file extension cannot answer either question.