At a glance
| Elapsed time | SLActive contact | SLA contact |
|---|---|---|
| 2 weeks | 14.8% | 12.2% |
| 4 weeks | 48.3% | 32.4% |
| 6 weeks | 61.6% | 61.5% |
A human study with planned retrieval
Lang and colleagues studied two implant surface treatments using small research devices in the back of the jaw. The devices were retrieved at prespecified intervals. This design allowed the team to examine newly formed bone directly, something a routine follow-up X-ray cannot show at that scale.


Why the time point changes the answer
At four weeks, the reported contact was 48.3% on the hydrophilic SLActive surface and 32.4% on SLA. By six weeks, values were 61.6% and 61.5%. Describing the four-week difference alone leaves out the later convergence observed in this experiment.
Where the evidence stops
This experiment used research devices and microscopic endpoints during early healing. It did not follow crowns under years of chewing. To evaluate a promise of earlier function, look for a clinical comparison that measured loading, complications and patient outcomes in the intended restoration and jaw.